Accuracy Analysis of Oxygen Saturation Measurements of TKDN Oximeters Using the Uncertainty Method

MEDIKA TRADA : Jurnal Teknik Elektomedik Polbitrada
Politeknik Bina Trada Semarang

📄 Abstract

This study aims to evaluate the accuracy and effectiveness of several domestic pulse oximeter brands (Elitech, Kaltech, and Endo) and a foreign brand (GE), calibrated using the Prosim SPOT Light SpO₂ Pulse Oximeter Tester. This quantitative study involved three domestically produced oximeters and one imported unit. The oxygen saturation setpoints were varied at 80%, 85%, 90%, 95%, 97%, 98%, 99%, and 100%, with three repetitions for each. Data analysis was conducted using the GUM Uncertainty Method and ANOVA statistical testing. The results show that all oximeters have combined uncertainty (expanded uncertainty, U₉₅) values between 0.58–3.23%, with reading deviations from reference values below ±2%, within acceptable tolerance limits. Kaltech and Endo oximeters demonstrated stable performance with no significant differences across saturation setpoints after calibration. Elitech and GE exhibited higher sensitivity to oxygen saturation variations, particularly at lower ranges. The One-Way ANOVA results indicated that Kaltech and Endo had p-values > 0.05, signifying no significant differences, while Elitech and GE showed p < 0.05, indicating differences between setpoints. Overall, domestically produced oximeters showed equivalent performance to imported devices after standard calibration, supporting the enhancement of Domestic Component Level (TKDN) and strengthening the independence of the national medical device industry.

🔖 Keywords

#Accuracy; Measurement Uncertainty; Pulse Oximeter; SPO2; Akurasi; Ketidakpastian Pengukuran; Pulse Oximeter; spo2

ℹ️ Informasi Publikasi

Tanggal Publikasi
04 December 2025
Volume / Nomor / Tahun
Volume 6, Nomor 2, Tahun 2025

📝 HOW TO CITE

Ari Widiatmoko; Ari Widiatmoko; Sri arttini Dwi prasetyowati, "Accuracy Analysis of Oxygen Saturation Measurements of TKDN Oximeters Using the Uncertainty Method," MEDIKA TRADA : Jurnal Teknik Elektomedik Polbitrada, vol. 6, no. 2, Dec. 2025.

ACM
ACS
APA
ABNT
Chicago
Harvard
IEEE
MLA
Turabian
Vancouver

🔗 Artikel Terkait dari Jurnal yang Sama

Pengembangan Digital Pressure Meter Berbasis IoT untuk Pengujian Tekanan Suction Pump dan Pencatatan Data Otomatis

Haris Kuspranoto, Abdul; Ulin Nuha ABA, Muhammad; Andika Hermawan, Fanhas

07 Aug 2026

DEVELOPMENT OF AN ULTRASOUND PHANTOM WITH MODIFIED VISCOSITY TO SIMULATE HUMAN SOFT TISSUE

Ulin Nuha ABA, M.Si, Muhammad; Haris Kuspranoto, Abdul; ., Mugiyanto; Hadi, Syaikhul; Satrio Anggono, Christopher Nicolaus; Ahmad ilham, Gus; Ulin Nuha ABA, M.Si, Muhammad; Haris Kuspranoto, Abdul; ., Mugiyanto; Hadi, Syaikhul; Satrio Anggono, Christopher Nicolaus; Ahmad ilham, Gus

29 Jun 2026

Analysis of the Influence Service Quality on Outpatient Satisfaction at Mojosongo Boyolali Community Health Center Moderated by Payment Type

Elsa Melva; Widiyoko, Aries; Wulandari, Sri; Elsa Melva; Widiyoko, Aries; Wulandari, Sri

27 Jun 2026

Analysis of Medical Records Staff Perceptions of RME Use at Indriati Solo Baru Hospital Using the TAM Model

AZZAHRA, EVA; Ratri Sukmaningsih, Wahyu; Ajika Pamungkas, Canggih; Ratri Sukmaningsih, Wahyu; Ajika Pamungkas, Canggih

27 Jun 2026

Analysis of the Implementation of Hospital Management Information Systems Using the HOT-Fit Method at Sebelas Maret University Hospital

LUTHFIYYAH, ANGGRAINI; Regina Putri, Frestiany; Ratri Sukmaningsih, Wahyu; Syarry Handoko, Anggraini Luthfiyyah; Regina Putri, Frestiany; Ratri Sukmaningsih, Wahyu

11 Jun 2026

Comparative Analysis of Manual and Electronic Medical Record Data Quality at Panti Waluyo Yakkum Hospital, Surakarta

Nurfika, Nurfika; Wulandari, Sri; Iskandar, Dwi; Nurfika, Nurfika; Wulandari, Sri; Iskandar, Dwi

10 Jun 2026

📊 Statistik Sitasi Jurnal