PRECISION AND ACCURACY ANALYSIS OF THE BIO SYSTEMS MODEL A15 CLINICAL CHEMISTRY ANALYZER IN GLUCOSE MEASUREMENT

MEDIKA TRADA : Jurnal Teknik Elektomedik Polbitrada
Politeknik Bina Trada Semarang

📄 Abstract

Blood glucose examination is one of the methods for diagnosing diabetes. Many people experience delays in glucose testing due to concerns about the accuracy of the testing equipment. To produce high-quality results, clinical laboratories require testing equipment with high accuracy and precision. This study aimed to evaluate the precision and accuracy of the A15 clinical chemistry analyzer in measuring glucose parameters. This was a quantitative descriptive study, involving 30 samples and calculating the precision using the coefficient of variation (CV) formula and accuracy using the d% formula. The results showed that the obtained CV value was 4.93%, below the maximum tolerance limit set by the Indonesian Minister of Health Regulation No. 43 of 2013, which is 5%. The obtained d% value was -0.036, indicating a high level of accuracy. It can be concluded that the A15 clinical chemistry analyzer has good precision and accuracy in measuring blood glucose.
 

🔖 Keywords

#Presisi #akurasi #Gluosa #koefisien Variasi; Precision; Accuracy; glucose; coefficient of variation

â„šī¸ Informasi Publikasi

Tanggal Publikasi
04 June 2025
Volume / Nomor / Tahun
Volume 6, Nomor 1, Tahun 2025

📝 HOW TO CITE

Marcelino, Al Vitto Marcelino; Nur, "PRECISION AND ACCURACY ANALYSIS OF THE BIO SYSTEMS MODEL A15 CLINICAL CHEMISTRY ANALYZER IN GLUCOSE MEASUREMENT," MEDIKA TRADA : Jurnal Teknik Elektomedik Polbitrada, vol. 6, no. 1, Jun. 2025.

ACM
ACS
APA
ABNT
Chicago
Harvard
IEEE
MLA
Turabian
Vancouver

🔗 Artikel Terkait dari Jurnal yang Sama

📊 Statistik Sitasi Jurnal